CS Instruments − DP 510

Portable dew point measurement down to −80 °Ctd with integrated data logger and one key advantage: an additional freely configurable input for external sensors (0–10 V, 4–20 mA, pulse, Modbus/RS-485).

Dew Point Measuring Range: −80…+50 °Ctd
Dew Point Accuracy: ±0.5 °Ctd (−10…+50 °Ctd), typ. ±2 °Ctd (outside range)
Data Logger: 16 GB (up to 100 million measurements)
Additional Input (DP 510): 0–10 V, (0) 4–20 mA, pulse, Modbus/RS-485
Data Transfer: USB (export via USB flash drive)
Pressure Range: −1…50 bar (standard); optional up to 350 bar
Battery: Li-Ion ~12 h; charging ~4 h
Display: 3.5" touchscreen display
Dew Point Measuring Range: −80…+50 °Ctd
Dew Point Accuracy: ±0.5 °Ctd (−10…+50 °Ctd), typ. ±2 °Ctd (outside range)
Data Logger: 16 GB (up to 100 million measurements)
Additional Input (DP 510): 0–10 V, (0) 4–20 mA, pulse, Modbus/RS-485
Data Transfer: USB (export via USB flash drive)
Pressure Range: −1…50 bar (standard); optional up to 350 bar
Battery: Li-Ion ~12 h; charging ~4 h
Display: 3.5" touchscreen display
Standard equipment 1 · Options and accessories 7
  • DP 510 – Portable Dew Point Meter Kit in Carrying Case (with Additional Input)The kit includes the DP 510 portable dew point meter with an additional input for external sensors, a portable measuring chamber up to 16 bar, a PTFE diffusion-tight hose with quick coupling (1 m length), power supply unit, 11.3% RH verification and calibration kit, quick shut-off coupling, dry storage container for sensor protection, and a large transport case (also suitable for additional sensors).
ParameterDP 510
Display3.5" touchscreen display
Dew Point Measuring Range [°Ctd]−80 … +50
Temperature Measuring Range [°C]−20 … +70
Relative Humidity Measuring Range [% RH]0 … 100
Dew Point Accuracy (−10 … +50 °Ctd) [°Ctd]± 0.5
Dew Point Accuracy (Outside Range) [°Ctd]typ. ±2 (extended range)
Calculated Moisture Parametersg/m³, mg/m³, ppm V/V, g/kg, °Ctdatm, % RH
Pressure Range (Standard) [bar]−1 … 50
Pressure Range (Special Version) [bar]−1 … 350
InterfaceUSB
Data Logger16 GB SD card (up to 100 million values)
Power Supply – Output Voltage [VDC]24 ±10%
Power Supply – Output Current [mA]120 (continuous operation)
Battery (Li-Ion) – Operating Time [h]approx. 12 (continuous operation)
Battery – Charging Time [h]approx. 4
Thread (Process Connection)NPT 1/2" or G 1/2" (stainless steel 1.4404)
Ambient Temperature [°C]0 … +50
EMCDIN EN 61326-1
Photo Key FunctionSaves the current screen as an image file (without additional software).
Additional Input for External Sensors (DP 510 only)1× freely configurable input for third-party sensors
Supported External Sensor Signal Types0–1/10 V, (0) 4…20 mA, pulse signal, Modbus/RS-485

Why Choose the DP 510?

Accurate Dew Point Measurement Down to −80 °Ctd

The DP 510 provides precise dew point measurement in compressed air and technical gas systems and enables reliable dryer diagnostics even at extremely low dew point values (down to −80 °Ctd).

Fast Response for Field Diagnostics

The fast response time enables rapid detection of dryer performance changes and immediate insight into process conditions during service inspections.

Touchscreen Display and Graphical Trend Curves

The 3.5-inch graphical touchscreen display provides intuitive operation, while measurement curves are displayed graphically, allowing dryer performance to be evaluated at a glance from the start of the measurement.

Integrated Data Logger: Up to 100 Million Measurements

The integrated 16 GB data logger allows storage of up to 100 million measurements. Logging intervals are freely configurable, and each measurement can be stored with a custom comment (e.g. measurement location).

Data Transfer via USB Flash Drive

Measurement data can be easily transferred to a computer using a USB flash drive. In addition, the 'Photo' function allows the current screen to be saved as an image file without additional software.

Automatic Calculation of All Key Moisture Parameters

The DP 510 automatically calculates important moisture parameters such as g/m³, mg/m³, ppm V/V, g/kg, % RH, and atmospheric pressure dew point, simplifying result interpretation in the field.

Additional Freely Configurable Input for External Sensors

The key advantage of the DP 510 is the second freely configurable input for third-party sensors. Supported signal types include 0–1/10 V, (0) 4–20 mA, pulse signals, and Modbus/RS-485. The DP 510 can also display values measured by external sensors.

Fast Connection and Complete Service Kit

Quick installation with a measuring chamber and quick coupling, combined with a dry storage container for sensor protection, ensures safe and efficient operation. The DP 510 is designed as an all-in-one solution for service teams, with everything organized in a transport case.

Compressed air dryer service

Dryer Service

Service & Diagnostics

Fast verification of refrigeration, membrane, and desiccant dryer performance through accurate on-site dew point measurements.

Maintenance of industrial compressed air systems

Industrial System Maintenance

Maintenance

A reliable tool for routine maintenance of compressed air and technical gas systems, enabling rapid detection of moisture-related issues and prevention of equipment failures.

Portable industrial dew point measurements

Portable Industrial Measurements

Field Measurements

The portable dew point meter with integrated data logging is ideal for field measurements, system commissioning, process validation, and temporary diagnostic inspections.

Measuring devices

Tell us what you measure and we will propose the right instrument.

Fill in a short form and we will prepare a proposal for measurement equipment and implementation based on the medium, measuring range, and measurement objective.

Proposal of the right instrument type and measuring range
Measurement reports, trends, and interpretation of results
Supply, installation, calibration, and service
Contact form
* required fields
Your message is sent directly to our email inbox.
Where to find us
Directions
Pod lipami 4 · 1370 Logatec, Slovenija

Dew Point Meters

Choose the right dew point sensor or meter for monitoring residual moisture in compressed air and technical gases. The portfolio includes fixed-installation probes, ATEX versions, and portable service instruments.