Dew point measurement

Dew point meters for compressed air and gases

The pressure dew point tells you how much moisture is still in the compressed air and whether the dryer really works. CS Instruments sensors and portable meters measure down to -80 °Ctd, raise an alarm when limits are exceeded and record the drying process.

measuring range
-80…+50 °Ctd
accuracy of the portable DP 500
± 0.5 °Ctd
special sensor versions
up to 500 bar
CS Instruments FA 500 dew point sensor with display

Why measure dew point

Dry air is the basis of a trouble-free system

When compressed air cools below its dew point, moisture separates as condensate. The results are pipe corrosion, failures of pneumatic controls and tools, freezing of outdoor lines and poorer product quality.

A refrigeration dryer cools the air to about +2 to +5 °C, an adsorption dryer dries it to -40 °Ctd or lower. The cooling temperature shown on the dryer is not enough – a faulty condensate drain, an open bypass or an overloaded dryer only show up in a dew point measurement.

  • Accurate down to -80 °CtdLong-term stable, condensation-resistant humidity sensors with a short response time.
  • Installation under pressureMeasuring chamber with quick coupling – fit or remove the sensor without a shutdown.
  • All humidity valuesBesides °Ctd also g/m³, mg/m³, ppm V/V, g/kg and °Ctd at atmospheric pressure.
  • Alarm and recordingRelays, 4…20 mA, Modbus-RTU or IO-Link for connection to a PLC and control system.

CS Instruments models

Dew point sensors and meters

Stationary sensors for continuous dryer monitoring, portable meters for service and sets with a display ready for immediate use.

StationaryCS Instruments FA 510 / FA 515

Stationary dew point sensor

FA 510 / FA 515

Pressure dew point sensor for refrigeration, membrane and adsorption dryers, with Modbus-RTU and IO-Link.

Measuring range
-80…+20 / -20…+50 °Ctd
Accuracy
± 1 °Ctd (+50…-20 °Ctd)
Pressure
-1…50 bar (500 bar)
Outputs
4…20 mA, Modbus-RTU, IO-Link
More about FA 510 / FA 515
StationaryCS Instruments FA 500

Sensor with display

FA 500

Dew point sensor with integrated display and alarm relay – ideal for all types of dryers.

Measuring range
-80…+20 / -60…+30 / -20…+50 °Ctd
Accuracy
± 1 °Ctd (+50…-20 °Ctd)
Pressure
-1…50 bar (500 bar)
Outputs
4…20 mA, Modbus-RTU, relay
More about FA 500
StationaryCS Instruments FA 550

Robust housing

FA 550

Sensor in a die-cast aluminium IP 67 housing for outdoor installation and harsh industrial environments.

Measuring range
-80…+20 / -60…+30 / -20…+50 °Ctd
Accuracy
± 1 °Ctd (+50…-20 °Ctd)
Pressure
-1…50 bar (350 / 500 bar)
Outputs
4…20 mA, Modbus-RTU, relay
More about FA 550
StationaryCS Instruments FA 515 Ex

ATEX version

FA 515 Ex

Dew point sensor for potentially explosive atmospheres and non-aggressive gases, up to 500 bar.

Measuring range
-80…+20 °Ctd
Accuracy
± 1 °Ctd (-20…+20 °Ctd)
Pressure
-1…500 bar
Outputs
4…20 mA, 2-wire
More about FA 515 Ex
StationaryCS Instruments FA 505

Compact sensor

FA 505

Long-term stable dew point sensor for OEM applications and adsorption dryers with little space.

Measuring range
-80…+20 °Ctd
Accuracy
± 1 °Ctd
Pressure
50 bar (350 bar)
Outputs
4…20 mA, Modbus-RTU
More about FA 505
PortableCS Instruments DP 500

Portable meter

DP 500

Service dew point meter down to -80 °Ctd with data logger and touch screen.

Measuring range
-80…+50 °Ctd
Accuracy
± 0.5 °Ctd (-10…+50 °Ctd)
Pressure
-1…50 bar (350 bar)
Memory
16 GB
More about DP 500
PortableCS Instruments DP 510

Portable meter

DP 510

Like the DP 500, with an additional free input for pressure, flow or a current clamp.

Measuring range
-80…+50 °Ctd
Accuracy
± 0.5 °Ctd (-10…+50 °Ctd)
Pressure
-1…50 bar (350 bar)
Additional input
4…20 mA, 0…10 V, pulse, Modbus
More about DP 510
PortableCS Instruments DP 400 mobile

Service case

DP 400 mobile

Mobile dew point meter with integrated pressure measurement in a sturdy case.

Measuring range
-80…+50 °Ctd
Accuracy
± 1 °Ctd (+50…-20 °Ctd)
Pressure
0…16 bar ± 0.5 %
Battery life
12 h
More about DP 400 mobile
StationaryCS Instruments DS 52

Dryer set

DS 52

DS 52 display with FA 510 sensor and two alarm relays for dryer monitoring.

Sensor inputs
1 × 4…20 mA
Display
LED, 5-digit
Alarm relays
2
Outputs
4…20 mA
More about DS 52
StationaryCS Instruments DS 400

2/4-input data logger

DS 400

Stationary touch-screen data logger for dew point, flow and pressure – with alarms and memory.

Sensor inputs
2 / 4
Display
3.5″ touch
Memory
100 million values
Alarm relays
2
More about DS 400

Comparison

Dew point meter comparison

ModelVersionMeasuring rangeAccuracyPressureOutputs
FA 510 / FA 515Stationary-80…+20 / -20…+50 °Ctd± 1 °Ctd (+50…-20 °Ctd)-1…50 bar (500 bar)4…20 mA, Modbus-RTU, IO-Link
FA 500Stationary-80…+20 / -60…+30 / -20…+50 °Ctd± 1 °Ctd (+50…-20 °Ctd)-1…50 bar (500 bar)4…20 mA, Modbus-RTU, relay
FA 550Stationary-80…+20 / -60…+30 / -20…+50 °Ctd± 1 °Ctd (+50…-20 °Ctd)-1…50 bar (350 / 500 bar)4…20 mA, Modbus-RTU, relay
FA 515 ExStationary-80…+20 °Ctd± 1 °Ctd (-20…+20 °Ctd)-1…500 bar4…20 mA, 2-wire
FA 505Stationary-80…+20 °Ctd± 1 °Ctd50 bar (350 bar)4…20 mA, Modbus-RTU
DP 500Portable-80…+50 °Ctd± 0.5 °Ctd (-10…+50 °Ctd)-1…50 bar (350 bar)–
DP 510Portable-80…+50 °Ctd± 0.5 °Ctd (-10…+50 °Ctd)-1…50 bar (350 bar)–
DP 400 mobilePortable-80…+50 °Ctd± 1 °Ctd (+50…-20 °Ctd)0…16 bar ± 0.5 %–
DS 52Stationary–––4…20 mA
DS 400Stationary–––USB, Ethernet, RS 485

Dryers and dew point

Which dew point to expect

Typical pressure dew points by dryer type and the matching ISO 8573-1 water classes.

+2…+5 °Ctd

Refrigeration dryer

Cools the air to 2–5 °C and separates the condensate. Corresponds to water class 4 (≤ +3 °C) or 5. Measurement reveals a faulty condensate drain and bypass.

-40 °Ctd

Adsorption dryer

The desiccant binds water vapour. Corresponds to class 2. Too short or too long regeneration and spent desiccant show up as a rising dew point.

-70 °Ctd

Strictest requirements

Class 1 for electronics, pharmaceuticals or outdoor lines in winter. Here we measure with a -80…+20 °Ctd range.

Granulate, gases

Granulate dryers and technical gases

The portable DP 500 checks plastic granulate dryers and residual moisture in N₂, O₂ and other gases.

ISO 8573-1:2010

Compressed air purity classes

ISO 8573-1 defines limit values for solid particles, water (pressure dew point) and oil. A class is written e.g. as ISO 8573-1 [1:2:1] – particles, water and oil in that order.

Water → dew point measurement

A refrigeration dryer reaches about +3 °Ctd (class 4), an adsorption dryer -40 °Ctd (class 2) or -70 °Ctd (class 1). Only a measurement shows whether the dryer really works.

Dew point meters →
ClassSolid particles – number per m³WaterOil (aerosol + vapour)
0.1–0.5 µm0.5–1 µm1–5 µmPressure dew pointmg/m³
0As specified by the user – stricter than class 1
1≤ 20 000≤ 400≤ 10≤ -70 °C≤ 0.01
2≤ 400 000≤ 6 000≤ 100≤ -40 °C≤ 0.1
3–≤ 90 000≤ 1 000≤ -20 °C≤ 1
4––≤ 10 000≤ +3 °C≤ 5
5––≤ 100 000≤ +7 °C–
6–––≤ +10 °C–

Source: CS Instruments catalogue 2025/26 (ISO 8573-1:2010).

How to choose

How to choose a dew point meter

  1. Dryer type

    For a refrigeration dryer choose the -20…+50 °Ctd range, for an adsorption dryer -80…+20 °Ctd.

  2. Continuous or occasional

    For continuous monitoring a stationary FA 510 or FA 500 sensor, for service and inspections the portable DP 500.

  3. Environment and pressure

    Outdoors or in dust FA 550 (IP 67), in Ex areas FA 515 Ex, above 50 bar the high-pressure version.

  4. Connection

    Stand-alone alarm (FA 500, DS 52) or connection via 4…20 mA, Modbus or IO-Link to a data logger.

FAQ

Dew point measurement questions

What is the pressure dew point?

The pressure dew point is the temperature at which moisture in compressed air starts to condense at operating pressure. It is given in °Ctd. The lower it is, the drier the air.

Where should the dew point sensor be installed?

Usually directly downstream of the dryer, where we check the dryer's performance. In more demanding systems we add measuring points after filters, at remote parts of the network and at critical consumers.

What is the difference between refrigeration and adsorption dryers for measurement?

A refrigeration dryer reaches about +3 °Ctd, so a sensor with a -20…+50 °Ctd range is sufficient. An adsorption dryer reaches -40 °Ctd or lower; for this choose a sensor with a -80…+20 °Ctd range.

Can the sensor be replaced or calibrated without a shutdown?

Yes. The sensor sits in a measuring chamber with a quick coupling, so it can be removed under pressure. A replacement sensor can be fitted during calibration.

How often should a dew point sensor be calibrated?

We recommend once a year. CS Instruments calibrates its own sensors and those of other manufacturers in the -80…+20 °Ctd range using DKD/BAM-certified references.

Measuring devices

Tell us what you measure and we will propose the right instrument.

Fill in a short form and we will prepare a proposal for measurement equipment and implementation based on the medium, measuring range, and measurement objective.

Proposal of the right instrument type and measuring range
Measurement reports, trends, and interpretation of results
Supply, installation, calibration, and service
Contact form
* required fields
Your message is sent directly to our email inbox.
Where to find us
Directions
Pod lipami 4 · 1370 Logatec, Slovenija